The device was not returned for analysis.
Since the device was not returned, we are unable to perform further root cause analysis.
All devices are 100% tested and all products are 100% inspected for damages and irregularities during manufacture.
If information is provided in the future, a supplemental report will be issued.
|
Medtronic received information that the pipeline flex device' pushwire broke.
The patient underwent embolization treatment for a unruptured amorphous supraclinoid internal carotid artery (ica), measuring 6mmx4mm, landing zone distal 3.
5mm proximal 3.
7mm.
The vessel was observed normal tortuous.
It was reported that catheter was flushed and prepared as per instructions for use (ifu).
It was delivered to the desired location without any trouble.
Ped-400-20 (ped) was advanced into the microcatheter.
It was pushed well without any resistance, it was then attempted to be deploying but the operator was not happy with the opening hence tried to resheathed it with the microcatheter.
Resistance was felt and all of a sudden pipeline pusher wire was moving freely into the catheter.
The complete system was taken out to inspect.
It was found that the distal portion of ped was broken and the catheter distal part was stretched.
No patient symptoms or complications were associated with this event.
Dual antiplatelet treatment was administered.
Pru level was 157.
The catheter was flushed as indicated per the ifu.
The post procedure angiography results show successful deployment of another ped.
|