• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 3 HOLE ACET SHELL 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Catalog Number 71335552
Device Problem Loose or Intermittent Connection (1371)
Patient Problem Injury (2348)
Event Date 11/07/2019
Event Type  Injury  
Event Description
It was reported that after tha on (b)(6) 2019 the patient was experiencing pain.A linear acetabular loosening was diagnosed.During the revision surgery, a hole change was performed.No patient injury or other complications were reported.
 
Manufacturer Narrative
The devices, used in treatment, were not returned for evaluation and the reported event could not be confirmed.The medical investigation concluded that this complaint from chili reports the revision of a hip (head, liner and shell) approximately 4 months post implantation, secondary to ¿pain¿ and loosening.No clinical/medical information has been provided for this investigation.Without supporting clinical/medical documents, a thorough investigation cannot be performed.Should information become available this complaint can be re-assessed.A review of complaint history did not reveal additional complaints for the listed batches.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.Pain is a potential complication associated with any surgery.Some potential probable causes of this event could include patient reaction or a post-operative healing issue.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 3 HOLE ACET SHELL 52MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9451741
MDR Text Key170277155
Report Number1020279-2019-04386
Device Sequence Number1
Product Code MBL
Combination Product (y/n)N
PMA/PMN Number
K070756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number71335552
Device Lot Number19CM13385
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-