Additional information was added to d10, h3, h4 and h6.H4: the potential lot number¿s were manufactured on: r19h14037- 08/16/2019, r19g22057- 07/23/2019, r19h07130- 08/08/2019, r19h23053- 08/24/2019.H10: the device was received for evaluation.All components are correctly placed and according to specifications.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed, including pressure and clear passage testing; and the device operated according to product specifications.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|