• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ALTRX NEUT 28IDX46OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ALTRX NEUT 28IDX46OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Catalog Number 122128046
Device Problem Device-Device Incompatibility (2919)
Patient Problem Not Applicable (3189)
Event Date 11/20/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: examination of the returned liner finds nothing outward to suggest product error.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Event Description
It was reported that the original implant date unknown.The surgeon was doing a total hip on this patient and implanted a bantam 46mm acetabular cup and implanted a 28mm neutral x 46mm pinnacle liner.He felt the liner did not set correctly.He impacted the liner a few times and then decided to remove the liner and insert a trial liner.The trial liner sat perfect in the cup.Another liner was opened and inserted into the cup without issue.I am returning the liner for inspection.Doe: (b)(6) 2019.Left hip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRX NEUT 28IDX46OD
Type of Device
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic drive
warsaw, IN 46581-0988
6107428552
MDR Report Key9457946
MDR Text Key188042213
Report Number1818910-2019-122676
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295015932
UDI-Public10603295015932
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K072963
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number122128046
Device Lot NumberJ3352T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/20/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/27/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age49 YR
-
-