• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS TSH ASSAY; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS TSH ASSAY; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE Back to Search Results
Model Number TSH
Device Problem Low Test Results (2458)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/01/2019
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).The sample was requested for investigation.
 
Event Description
The initial reporter complained that results for 1 patient tested for elecsys ft3 iii (ft3 iii), elecsys ft4 iii (ft4 iii) and elecsys tsh (tsh) on a cobas e 801 module do not fit the patient's clinical picture.The ft4 iii were discrepant when compared to the abbott method.The date of event is an approximation as the date of event was unknown.This medwatch will cover tsh.Refer to medwatch with patient identifier (b)(6) for information on the ft3 iii results and medwatch with a1 patient identifier (b)(6) for information on the ft4 iii results.The e801 module serial number was (b)(4).
 
Manufacturer Narrative
The sample was received for investigation.No interfering factor was identified.Assays from different manufacturers can generate different results.This relates to the overall setups of the assays, the antibodies used, differences in reference materials and the standardization methodology used.The investigation did not identify a product problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TSH ASSAY
Type of Device
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key9465526
MDR Text Key219765565
Report Number1823260-2019-04404
Device Sequence Number1
Product Code JLW
Combination Product (y/n)N
PMA/PMN Number
K961491
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 02/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTSH
Device Catalogue Number08443432190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Date Manufacturer Received11/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CARBIMAZOL
Patient Age60 YR
-
-