Model Number 97713
Device Problems Failure to Deliver Energy (1211); Material Frayed (1262); Charging Problem (2892); Connection Problem (2900); Material Integrity Problem (2978)
Patient Problems Pain (1994); Therapeutic Response, Decreased (2271)
Event Date 11/05/2019
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 37761, serial#: (b)(4), product type recharger:, other relevant device(s) are: product id: 37761, serial/lot #: (b)(4), analysis of the recharger c0612743 found evidence of a frayed cord.Fdc code 4315 applies to the recharger.Fdc code 67 applies to the implantable neurostimulator.Fdm 10 and fdr 140 apply to the recharger.Fdm 4117 and fdr 3221 apply to the implantable neurostimulator.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received regarding a patient with an implantable neurostimulator (ins), and it was reported that the recharger would not charge from the wall.The patient stated the desktop charger cord was frayed, and the connector pin loose.The patient noted the ins would not charge and the reposition antenna screen was seen.The patient had not been able to charge for a month so she took aspirin for her sciatic nerve pain, because she did not have therapy, since she could not charge.The patient had not been able to charge for weeks.The patient did not have a patient programmer.Patient was issued a new desktop charger and recharger and it was reviewed that she should follow up with her healthcare provider (hcp) if these did not resolve the issue.There were no reported complications and no further complications were expected.Patient was implanted for spinal pain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key9468942
MDR Text Key176997864
Report Number3004209178-2019-23805
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169109490
UDI-Public00643169109490
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/14/2014
Device Model Number97713
Device Catalogue Number97713
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2019
Date Manufacturer Received11/26/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/20/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
-
-