MEDTRONIC SOFAMOR DANEK USA, INC CD HORIZON BALLAST SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
|
Back to Search Results |
|
Catalog Number 25840018590 |
Device Problem
Migration or Expulsion of Device (1395)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
Injury
|
Manufacturer Narrative
|
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the r eported event.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient underwent open lumbar fusion at l2-pelvis.On an unknown date, post-op, set screw dislodged from the pedicle screw.Hence, a revision surgery was performed, in which the loose screw was removed and was replaced with bigger screw.No other patient complications were reported.
|
|
Manufacturer Narrative
|
X-ray review results: post-op x-rays for multilevel thoracic to pelvic fixation were provided.Interbody grafts are present at l5-s1, l4-l5 and l3-l4 appear to be previous artificial disc replacements.One of the set screws at the sacrum is out of the screw head.Fusion status is unknown.If information is provided in the future, a supplemental report will be issued.
|
|