RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Device Alarm System (1012)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2019.Date of report: 12/16/2019.
|
|
Event Description
|
The customer reported alarm (led) light emitting diode error.It is unknown if the ventilator was being used on a patient at the time that the error was discovered.Confirmation has been requested.The manufacturer¿s field service engineer (fse) confirmed the reported alarm led failure issue.The manufacturer¿s field service engineer (fse) remotely confirmed with the customer that the power management board was replaced, but was still receiving the alarm led failure.The fse advised to replace the power switch overlay, and the customer confirmed the power switch overlay fixed the issue.
|
|