• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE Back to Search Results
Model Number LTF-S190-5
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/19/2019
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus medical systems corp.(omsc) for evaluation.Omsc confirmed that the endoscopic image on the monitor was disappeared and error b30 was displayed on the monitor after a long period of applying the power.Therefore it was consider that some electrical component of the subject device was damaged.Also, the subjected device had been used for seven years without any repair.The exact cause of the reported phenomenon could not be conclusively determined, however there was the possibility that the electrical component damage was attributed to deterioration due to the long-term use.Also omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.If significant additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that the endoscopic image of the subject ltf-s190-5 disappeared during an unspecified therapeutic procedure.The user replaced the subject device to another unspecified device to complete the procedure.The user checked with taking time after the procedure, but the phenomenon which is disappearing the image was not duplicated.There was no report of the patient injury other than replacing the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key9479056
MDR Text Key211732354
Report Number8010047-2019-04377
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K080948
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-S190-5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2019
Was the Report Sent to FDA? No
Date Manufacturer Received11/27/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/03/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-