OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P6R
Device Problem Material Protrusion/Extrusion (2979)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to the service center for evaluation.The user¿s complaint was confirmed.A visual inspection was performed on the received condition and noted the bending section skeleton tab was protruding (sticking out) from the bending section cover near the insertion tube area.The bending section cover was removed and the bending section skeleton is fully broken and separated with no signs of any sharp edges.Excessive broken fibers were noted on the image due to the broken bending section skeleton.The instruction manual states the following; ¿do not twist or bend the bending section with your hands¿.Additionally, the instructions for safe use manual indicate that, damage to the bending section would happen if excessive force was applied while angulating in the opposite direction if there was no movement from the bending section; this would occur more so in a narrow environment.Based on the investigation, the cause of the broken bending section skeleton is due to mishandling.
 
Event Description
The service center was informed that the scope¿s tip was noted to be bent.There was no patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key9480714
MDR Text Key219764409
Report Number8010047-2019-04382
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P6R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2019
Was the Report Sent to FDA? No
Date Manufacturer Received11/22/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-