• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. XPS® BLADE; BUR, EAR, NOSE AND THROAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. XPS® BLADE; BUR, EAR, NOSE AND THROAT Back to Search Results
Model Number 1884012HR
Device Problems Material Separation (1562); Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/25/2019
Event Type  malfunction  
Manufacturer Narrative
Analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Event Description
A healthcare professional (hcp) reported via manufacturer representative that when they attached the blade to the handpiece during the procedure, the blade stopped running and broke into two segments.There were no intervention planned or performed.There were no broken pieces remained inside the patients body.The procedure was completed with backup product(s).There was no patient impact.
 
Manufacturer Narrative
H3: analysis found that the inner shaft broke 0.53¿ from the distal face of the inner hub which would have resulted in the reported malfunction.There were striations around the outside diameter of the break point indicating metal on metal contact during use.The break point corresponds to the proximal end of the outer tube in the front hub.There was no damage to the distal tip.When viewed under magnification, there was deformation and indentations of the front hub locking area consistent with aggressive use.There were no bends or signs of a concentricity issues.H6: additional information suggests that fdm: 4118,fdr:3233 and fdc:11 no longer apply.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® BLADE
Type of Device
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key9488329
MDR Text Key173373598
Report Number3004209178-2019-24035
Device Sequence Number1
Product Code EQJ
UDI-Device Identifier00885074080604
UDI-Public00885074080604
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/20/2023
Device Model Number1884012HR
Device Catalogue Number1884012HR
Device Lot NumberHG39BWP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2019
Date Manufacturer Received01/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-