Catalog Number 4845-4-106
Device Problem Loss of Osseointegration (2408)
Patient Problems Pain (1994); Injury (2348); Inadequate Osseointegration (2646)
Event Date 11/18/2019
Event Type  Injury  
Manufacturer Narrative
It was noted that the device is not available for evaluation. If additional information is received, it will be provided in a supplemental report upon completion of the investigation. A voluntary recall ra 2012-067 was initiated for abgii and rejuvenate modular stems and necks due to the potential risks associated with these devices. Not returned.
 
Event Description
As reported: "we removed a trident cluster cup & abg modular stem yesterday. Failed components: trident cluster cup & abg stem. Adverse consequences: groin and thigh pain. I don't have access to medical records due to potential hipaa violations. I did get a picture of the x-ray the pa brought to surgery. " update: "no known allegations against the liner, biolox head, or dome hole plug. Right side, intraop findings were as follows: the trident shell was suspected to be anteverted too much possibly causing iliopsoas pain and the abg stem was causing thigh pain possibly due to the stem being loose".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameABGII. MODULAR STEM
Type of DevicePROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
Manufacturer Contact
brad curtis
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key9488907
MDR Text Key189341133
Report Number0002249697-2019-04047
Device Sequence Number1
Product Code MEH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092406
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation
Remedial Action Recall
Type of Report Initial
Report Date 12/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Expiration Date10/31/2015
Device Catalogue Number4845-4-106
Device Lot NumberG2999716
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/18/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/27/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-2089-2012

Patient Treatment Data
Date Received: 12/18/2019 Patient Sequence Number: 1
-
-