Additional device product code: hrs.Complainant part is not expected to be returned for manufacturer review/investigation.A device history record (dhr) review was conducted: part: 04.211.050s, lot: 6l09487, manufacturing site: selzach, supplier: (b)(4), release to warehouse date: 16 sept 2019, expiry date: 01 sept 2029.Since there is no allegation against packing or sterility dhr review is done for non-sterile part: part number: 04.211.050, lot number: 14l8700, manufacturing site: monument, release to warehouse date: 28 august 2019, manufacturing location: monument, manufacturing date: 14-aug-2019, part number: 04.211.050, 2.7mm ti va lckng scr slf-tpng with t8 stardrive recess 50mm, lot number: 14l8700 (non-sterile), lot quantity: 151.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Component part(s) reviewed: part number: 04.211.050.999, 2.8mm ti screw blank 50mm 02.7 variable angle w/sd8, lot number: h754676, lot quantity: 921.Work order traveler met all inspection acceptance criteria.Inspection sheet, inspect warm head blank/inspect turn head and point, ns072130 rev b met all inspection acceptance criteria.Part number: 23032, tialnbci2.78, lot number: h574287, lot quantity: 603 lbs.Non-conformance was initiated against this lot for missing information on the supplier certification (missing ship date).The disposition of the nr was ¿rework¿ and a corrected certification was supplied.Certified test report (corrected) supplied by (b)(4) dated 06-feb-2018 was reviewed and determined to be conforming.Lot summary report dated 09-mar-2018 met all inspection acceptance criteria.Raw material receiving/putaway checklist met all inspection acceptance criteria.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2019, the patient underwent an open reduction internal fixation surgery for the proximal ulna fracture with the screwdriver in question.During the surgery, the surgeon could not insert the locking screw into the bone completely because the head of the locking screws were getting worn-out.The torque limitation mechanism seemed not to be working properly.Both the surgeon and a supervising physician tried to insert very carefully, but the problem didn¿t improve.The surgery was completed with the screw inserted incompletely.It is unknown out of six screws which screw was inserted incompletely.The surgery was delayed by less than thirty (30) minutes.The surgeon commented that there is a possibility that either the connection of the screwdriver and the screw was incomplete or the torque limiting attachment was defective.The surgeon worries that the screw may come out because of the incomplete insertion.No further information is available.Concomitant device reported: unknown plate (part# unknown ,lot# unknown, quantity 1).This report is for one (1) 2.7mm ti va locking screw 50mm.This is report 4 of 7 for complaint (b)(4).
|