Age/date of birth: unknown/ not provided.Sex/gender: unknown/ not provided.Expiration date: unknown as lot number was not provided.Lot number: unknown/not provided.Unique identifier (udi#): unknown as lot number not provided.Catalog number: a complete catalog number is unknown as lot number not provided.If implanted; give date: n/a (not applicable).The healon gv pro is not an implantable device.If explanted; give date: n/a (not applicable).The healon gv pro is not an implantable device.Facility name: unknown/ not provided.Phone number: unknown/ not provided.Address: unknown/ not provided.Device manufacture date: unknown, as the lot number was not provided.This complaint is part of the recall - report number 2020664-12/02/19-001-r: johnson & johnson surgical vision (jjsv) issued a voluntary recall on november 22, 2019.It has been reported customers have described healon gv pro as behaving differently than the legacy healon gv, especially in regard to the techniques required to remove the product from the eye.An increase of intra-ocular pressure (iop) is reported if there are small amounts of healon gv pro remaining behind the operative eye.This voluntary recall is being initiated due to received reports of healon gv pro being difficult to remove from the eye, leading to increased post-operative iop requiring additional intervention.Potential clogging of phacoemulsification equipment tubing has also been reported, which may lead to delay in the procedure or ocular injury.There are twenty-one affected lot numbers.The recall notification letter has been sent to all customers instructing them to return the units of healon gv pro from the twenty-one (21) affected lots.Johnson & johnson surgical vision has initiated a corrective and preventative actions (capa) to investigate and address the issue.Action items generated from the capa will be submitted in the future interim report(s) as part of the recall process.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
It was reported that a doctor is upset because of thicker, more dense viscoelastic.It was stated that the healon gv pro gets stuck behind the pieces and builds pressure in the eye creating a dangerous degree of increase in intraocular pressure.Additionally, it also tends to remain stuck behind the intraocular lens (iol) and will not come out of the eye easily after the iol is inserted.No other information is provided.
|