• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US CLEARCANNULA-THREADED 8.5MMX75MM 5PK -ST; ARTHROSCOPIC DRAINAGE CANNULA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US CLEARCANNULA-THREADED 8.5MMX75MM 5PK -ST; ARTHROSCOPIC DRAINAGE CANNULA Back to Search Results
Model Number 214120
Device Problem Detachment of Device or Device Component (2907)
Patient Problems No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Event Date 12/05/2019
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported by the affiliate in (b)(6) that during a shoulder arthroscopic surgical procedure, it was observed that the clear washer on the clearcannula-threaded 8.5mmx75mm 5pk -st device came apart from the cannula and ended up in patient's shoulder.According to the report, the device was being used as a portal into the shoulder to be able to use the arthroscopic instruments to be inserted.This was noted by the scrub team and the washer was able to be removed with no consequences.This was completed as a "near miss".There was no delay in the surgical procedure.There was a spare device available for use to complete the surgery.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: additional information: h6: method codes: device history lot
=
> a manufacturing record evaluation was performed for the finished device lot number 1902039, and no non-conformances were identified.Investigation summary
=
> the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.A manufacturing record evaluation was performed for the finished device lot number 1902039, and no non-conformances were identified.Since the complaint device was discarded, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARCANNULA-THREADED 8.5MMX75MM 5PK -ST
Type of Device
ARTHROSCOPIC DRAINAGE CANNULA
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9497638
MDR Text Key208153435
Report Number1221934-2019-60066
Device Sequence Number1
Product Code GEA
UDI-Device Identifier10886705003617
UDI-Public10886705003617
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2024
Device Model Number214120
Device Catalogue Number214120
Device Lot Number1902039
Was Device Available for Evaluation? No
Date Manufacturer Received01/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-