• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MILACA, INC. DUAL CHAMBER TEMPORARY PACEMAKER; PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MILACA, INC. DUAL CHAMBER TEMPORARY PACEMAKER; PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE Back to Search Results
Model Number 5388
Device Problem Inadequate User Interface (2958)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/10/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generator (epg) needed repair.The epg was returned for service.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product analysis: analysis confirmed that the external pulse generator (epg) needed repair.Analysis noted that the incoming test for the heart wire connector b atrial heart wire failed, the output connector was broken, and there were missing rings, ring cover, bails, and bail covers.The device was sent back to the customer unrepaired as the necessary parts were not available due to the model no longer being serviced by the company.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL CHAMBER TEMPORARY PACEMAKER
Type of Device
PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC MILACA, INC.
900 sixth avenue ne
milaca MN 56353 3728
Manufacturer (Section G)
MEDTRONIC MILACA, INC.
900 sixth avenue ne
milaca MN 56353 3728
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9498244
MDR Text Key179301716
Report Number2183613-2019-00223
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
P820003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5388
Device Catalogue Number5388
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/13/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/21/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-