BARD PERIPHERAL VASCULAR, INC. ATLAS PTA DILATATION CATHETER PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number AT75144
Device Problems Break (1069); Entrapment of Device (1212); Retraction Problem (1536)
Patient Problems Death (1802); Hematoma (1884); Thrombosis (2100)
Event Date 11/22/2019
Event Type  Death  
Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed. The sample was not returned to the manufacturer for inspection/evaluation. Therefore, the investigation of the reported event is inconclusive. Based upon the available information, the definitive root cause for this event is unknown. The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device. Upon receipt of new or additional information, a follow-up report will be submitted as applicable. (expiry date: 01/2022).
 
Event Description
It was reported that during an angioplasty procedure in the subclavian vein, after the device had been deflated, the pta balloon catheter was allegedly stuck in the vessel and was unable to be removed through the 7fr introducer sheath. Therefore, the introducer sheath was removed and an additional unsuccessful attempt was made to remove the catheter from the patient. A scalpel was then used to release pressure from the balloon. The balloon was again pulled until it was finally released, causing blood loss; pressure was applied and the graft was compromised, as it became fully clotted. Reportedly, the patient was transported and stable post-surgery; however, the patient expired the same night.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameATLAS PTA DILATATION CATHETER
Type of DevicePTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
Manufacturer Contact
judith ludwig
1415 w. 3rd street
tempe, AZ 85281
4803032689
MDR Report Key9499163
MDR Text Key172184175
Report Number2020394-2019-05701
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120971
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberAT75144
Device Catalogue NumberAT75144
Device Lot NumberGFDN2808
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/29/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 12/19/2019 Patient Sequence Number: 1
-
-