• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION COYOTE CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION COYOTE CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24698
Device Problems Material Rupture (1546); Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/12/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
It was reported that balloon rupture occurred. The 100% stenosed target lesion was located in the moderately tortuous and severely calcified superficial femoral artery. A jupiter 45 guidewire was advanced but failed to cross. After a non-bsc guidewire crossed the lesion, a 3. 0mm x 150mm x 150cm coyote balloon catheter was advanced but also failed to cross. An unknown 1. 5mm balloon catheter was advanced for dilatation, followed by the same coyote balloon that failed cross initially. However, during the first inflation, the balloon ruptured. A non-bsc stent was placed and the procedure was completed after confirming that vascular patency was good. No patient complications nor injuries were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCOYOTE
Type of DeviceCATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9501339
MDR Text Key176769656
Report Number2134265-2019-16033
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111295
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial
Report Date 12/19/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/20/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date04/01/2021
Device Model Number24698
Device Catalogue Number24698
Device Lot Number0021934513
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/13/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/02/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-