Model Number 5391
Device Problem No Pacing (3268)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/13/2019
Event Type  malfunction  
Manufacturer Narrative
This report is related to initial report from importer oscor with importer report # (b)(4).The device was returned to manufacturer and arrived on 28oct2019.Initial evaluation did not show any functional problem.Only the battery compartment lid was missing.The device underwent 48h long time testing in temperature cycling cabinet.No missing beat has been detected.The evaluation did not raise any function defect causing stimulation problems.The failure description could not be reproduced.The pacemaker works according to its specification.Possible causes for stimulation failures could be: damaged extension or adapter cable (was not returned for evaluation), bad lead to tissue contact, unappropriated lead position, connection problem between lead wire to patient extension cable or extension cable to pacemaker, not correct adjustment of capture threshold due to wrong stimulation amplitude setting.Unfortunately there were no further details of the circumstances available which could explain the most likely root cause.The missing battery compartment lid has been replaced and after calibration and completed safety and functional test returned to the importer on 05dec2019.
 
Event Description
It was reported that the external pulse generator (epg) did not stimulate.The epg was returned for service.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC SINGLE CHAMBER EXTERNAL PACEMAKER
Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
OSYPKA MEDICAL GMBH
albert-einstein-strasse 3
berlin, berlin 12489
GM  12489
Manufacturer (Section G)
OSYPKA MEDICAL GMBH
albert-einstein-strasse 3
berlin, berlin 12489
GM   12489
Manufacturer Contact
thilo thuemecke
albert-einstein-strasse 3
berlin, berlin 12489
GM   12489
MDR Report Key9503005
MDR Text Key220057590
Report Number9681449-2019-00005
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K033130
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5391
Device Catalogue Number5391
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2019
Date Manufacturer Received10/28/2019
Was Device Evaluated by Manufacturer? Yes
Type of Device Usage Reuse
Patient Sequence Number1
-
-