Model Number 254835
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/05/2019
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.Udi: (b)(4).The lot number is not currently available.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported by the sales rep that post-op an anterior cruciate ligament (acl) procedure, while cleaning the inner shaft of the extra large 30 mm sheath inserter, it was observed that a wire brush was stuck inside.There was no delay in the surgical procedure.There was a spare device available for use to complete the surgery.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary
=
> according to the information provided, it was reported that during while cleaning the inner shaft a wire brush was stuck inside.  the complaint device was received and evaluated.Visual observation reveals the distal end of the inserted looks fairly new but in used condition.The tip is slightly scratched, and evaluating it could not find any wire brush and without the foreign material inside the unit cannot verified.The possible root cause for the reported failure cannot be determined.As a potential cause cannot be associated to manufacturing, therefore a manufacturing record evaluation is not required.At this time, no corrective action is required, and no further action is warranted, as the device is fully functional.However, depuy synthes mitek will continue to track any related complaints within this device family to monitor the extent to which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRFX ADV - XL 30MM INSERTER
Type of Device
ORTHOPAEDIC IMPLANT INSERTER/EXTRACTOR
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key9503460
MDR Text Key205267012
Report Number1221934-2019-60074
Device Sequence Number1
Product Code LXH
UDI-Device Identifier10886705026357
UDI-Public10886705026357
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number254835
Device Catalogue Number254835
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2020
Date Manufacturer Received01/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-