MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 35 PTA CATHETER PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 35 PTA CATHETER PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number B2070-060
Device Problems Leak/Splash (1354); Material Rupture (1546)
Patient Problem No Patient Involvement (2645)
Event Date 12/06/2019
Event Type  Malfunction  
Manufacturer Narrative

Exemption number e2019001. The customer reported the device was discarded. Investigation is not yet complete. A follow-up report will be submitted with all additional relevant information.

 
Event Description

It was reported that during preparation of the balloon catheter when pulling negative air bubbles were seen. Therefore, the balloon catheter was decided not to be used. The balloon was pressurized by the technician and it was noted the balloon had a pinhole. A new balloon catheter was used to complete the procedure. There was no patient involvement and no clinically significant delay in the procedure. No additional information was provided.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARMADA 35 PTA CATHETER
Type of DevicePERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key9506007
MDR Text Key178336327
Report Number2024168-2019-14861
Device Sequence Number1
Product Code LIT
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK111899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,HEALTH
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/15/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received12/20/2019
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue NumberB2070-060
Device LOT Number90627G1
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/03/2020
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured06/27/2019
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-