• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE MEDIAL CONGRUENT (MC) RIGHT 10 MM; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE MEDIAL CONGRUENT (MC) RIGHT 10 MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problems Positioning Failure (1158); Difficult to Insert (1316)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/03/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.(b)(4).Concomitant medical product(s): item # 42522100711 lot # 63857827 unknown tibial tray.Report source: foreign: (b)(6).Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2019-00356.
 
Event Description
It was reported that during a total knee arthroplasty the surgeon was unable to lock persona mc 10mm surface to tibial tray.So the surgeon attempted to use a 11mm of persona mc surface but the same issue occurred again.Subsequently, the surgeon was able to complete the procedure with a cr 10mm surface.Attempts have made and no further information has been provided.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d10, g4, g7, h1, h2, h3, h6, h10.This complaint was not confirmed.Visual examination of returned products found that both articular surfaces show signs of use (nicks or gouges) and the dovetail feature is flared.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE MEDIAL CONGRUENT (MC) RIGHT 10 MM
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
MDR Report Key9510699
MDR Text Key190683855
Report Number3007963827-2019-00354
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
PMA/PMN Number
K150090
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 04/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42522100710
Device Lot Number64417075
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2020
Was the Report Sent to FDA? No
Date Manufacturer Received04/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-