(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.(b)(4).Concomitant medical product(s): item # 42522100711 lot # 63857827 unknown tibial tray.Report source: foreign: (b)(6).Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2019-00356.
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d10, g4, g7, h1, h2, h3, h6, h10.This complaint was not confirmed.Visual examination of returned products found that both articular surfaces show signs of use (nicks or gouges) and the dovetail feature is flared.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|