Model Number 3ZZ*FX25RWCA
Device Problem Use of Device Problem (1670)
Patient Problems Injury (2348); Blood Loss (2597)
Event Date 12/12/2019
Event Type  Injury  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed. Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available. (b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, the oxygenator would not oxygenate. The patient was under anesthesia, ventilating and came off bypass. The vaporizer was bypassed and bypass initiation was tried again and was unsuccessful. The oxygenator was then changed out at a slow pace. There was a delay for 40 minutes, there was a blood loss of 200ml, the product was changed out, the surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNS FX25RWC W/RES
Type of DeviceBLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
cathleen hargreaves
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key9512215
MDR Text Key188527154
Report Number1124841-2019-00346
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 04/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date02/28/2022
Device Model Number3ZZ*FX25RWCA
Device Catalogue NumberN/A
Device Lot NumberXD18
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received04/16/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 12/23/2019 Patient Sequence Number: 1
-
-