Model Number 8401
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Evaluation results: one capnochk ii was returned for investigation.The reported complaint "co2 sensor failure" was verified.The monitor was powered up and the monitor had no flowrate, then the monitor was opened up and the valve elbow was broken off in the naphion tubing due to impact suffered to the monitor.The investigator replaced the valve and naphion tubing and now the monitor has flowrate of 109 spec is 120+/-20ml/min so a restrictor was replaced to bring the flow up above the nominal readings of 133 ml/min.A factory calibration was then performed and it passed.
 
Event Description
It was reported that the device exhibited a co2 sensor failure.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPNOCHK II
Type of Device
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC
3350 granada avenue north
suite 100
oakdale MN 55128
Manufacturer Contact
dave halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key9513415
MDR Text Key176883301
Report Number3012307300-2019-07590
Device Sequence Number1
Product Code CCK
UDI-Device Identifier10610586036781
UDI-Public10610586036781
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K991086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number8401
Device Catalogue Number8401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2019
Date Manufacturer Received11/27/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/23/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-