• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC DRIVE SHAFT-MINIMUM 520MM LENGTH-FOR USE WITH RIA; REAMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC DRIVE SHAFT-MINIMUM 520MM LENGTH-FOR USE WITH RIA; REAMER Back to Search Results
Model Number 314.743
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 01/01/2019
Event Type  malfunction  
Manufacturer Narrative
Exact date of event is unknown.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is synthes sales consultant.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on an unknown date, a reamer/irrigator/aspirator (ria) drive shaft was being disassembled by sterile processing department (spd) and when trying to remove an unknown ria reamer head from drive shaft the tongs of the drive shaft were mangled.Spd tried using pliers to remove ria head with twisting motion and mangled/broke the ria driveshaft connection.There was no patient involvement.Unknown pliers (part# unknown, lot# unknown, quantity 1).This report is for one (1) drive shaft-minimum 520mm length-for use with ria.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: part: 314.743 synthes lot: h220887, supplier lot: h220887, release to warehouse date: may 16, 2017; june 05, 2017.Supplier: criterion tool & die, inc.There are no relevant nonconformance reports.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Visual inspection: the drive shaft-minimum 520mm length-for use with ria was received at us customer quality (cq).Visual inspection of the complaint device showed the distal tip of the device had a fragment broken off.Dimensional inspection: no dimensional inspection can be performed due to post manufacturing damage.Document/specification review: current and manufactured versions were reviewed.No design issues or discrepancies were identified.Investigation conclusion: this complaint is confirmed as the distal tip of the complaint device has a fragment broken off.No definitive root cause could be determined based on the provided information.No new, unique, or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: a review of the device history record has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRIVE SHAFT-MINIMUM 520MM LENGTH-FOR USE WITH RIA
Type of Device
REAMER
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key9517101
MDR Text Key190648858
Report Number2939274-2019-62778
Device Sequence Number1
Product Code HTO
UDI-Device Identifier10886982189042
UDI-Public(01)10886982189042
Combination Product (y/n)N
PMA/PMN Number
K042899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 12/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/24/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number314.743
Device Catalogue Number314.743
Device Lot NumberH220887
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2020
Date Manufacturer Received01/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNK - PLIERS
-
-