Model Number 1000
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Patient Involvement (2645)
Event Date 12/03/2019
Event Type  malfunction  
Manufacturer Narrative
Physio-control evaluated the customers device and verified the reported issue. The customer was provided a replacement device. Physio-control further evaluated the customers device and observed that integrated circuit designator u1 on the therapy pcb assembly has been compressed on one end, and the other end has risen a leg off the pcb ferrite bead and is pressing on integrated circuit designator u1, this was determined to be the cause of the reported issue. The device was destructively tested.
 
Event Description
The customer contacted physio-control to report that their device delivered monophasic shock. In this state the device may deliver an inappropriate defibrillation therapy if needed. There was no patient use reported with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameLIFEPAK® 1000 DEFIBRILLATOR
Type of DeviceAUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key9520090
MDR Text Key190672028
Report Number0003015876-2019-02069
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122600
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation
Type of Report Initial
Report Date 12/26/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number1000
Device Catalogue Number99425-000025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2019
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/03/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/05/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-