• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL FMS VUE PUMP; DISTENSION UNIT, FLUID, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL FMS VUE PUMP; DISTENSION UNIT, FLUID, ARTHROSCOPIC Back to Search Results
Catalog Number 284002
Device Problem Increase in Pressure (1491)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/10/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).Device was used for treatment, not diagnosis.Udi: (b)(4).
 
Event Description
It was reported by the customer via phone that the fms vue pump is making a loud noise and flow rate is unsteady.No surgery information available.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary
=
> the complaint device was received at the service center and evaluated.The customer reported an issue of the device was making a loud noise and flow rate was unsteady.Per service manual operational and diagnostic, this complaint can be confirmed.During evaluation, worn gear was identified.It was replaced to resolve the issue.After repair, the device was found to be working according to the specifications.The worn gear is identified as the root cause of the reported problem.The service history has been reviewed in lieu of the device history record for this device since it was previously serviced.The device was last serviced on 03/06/2019 and passed all functional testing before being returned to the customer.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot
=
> the service history has been reviewed in lieu of the device history record for this device since it was previously serviced.The device was last serviced on 03/06/2019 and passed all functional testing before being returned to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS VUE PUMP
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key9525605
MDR Text Key205993483
Report Number1221934-2019-60174
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705020591
UDI-Public10886705020591
Combination Product (y/n)N
PMA/PMN Number
K130169
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 12/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number284002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2019
Date Manufacturer Received01/12/2020
Patient Sequence Number1
-
-