Catalog Number 2400L |
Device Problem
Break (1069)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/04/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a load cell module had frayed cables; further described as "cords exposed and some metal sticking out".This was identified during an unspecified process step.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
The device was received for evaluation.A visual inspection was performed which observed the outer insulation covering of the cable was damaged/frayed exposing the ground braid and wires.No actual internal conductors exposed.Functional testing was performed which could not establish load cell communication; there was intermittent connectivity or continuity of signal wires due to the damaged frayed cord.The reported condition was verified.The cause of the condition was due to the electrical component failure.The scale cable assembly was replaced during the service refurbishment process.A service history review revealed no indication that the parts replaced caused or contributed to the reported event; the device was serviced greater than twelve (12) months.Should additional relevant information become available, a supplemental report will be submitted.
|
|