Model Number TABLETOP-JAPAN
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that the aspiration failed during a procedure.The product was replaced and procedure completed with no patient harm.
 
Manufacturer Narrative
Additional information provided in d.10., g.1., g.2., h.3., h.6., and h.10.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The used returned sample was visually inspected and no obvious defects were found.A console representing the current software version was used to test the sample.The ball in the drip chamber¿s check valve moved freely per specification.The sample could prime and tune with the ultrasonic handpiece successfully.The irrigation and aspiration pressure were measured at multiple set points and met specifications.No message code appeared on the screen.Fluid flowed from balanced salt solution (bss) to the drain bag without any interfering.No leakage was detected from the pump elastomer or on the pump area of the fluidics module.Cleaning process was able to be performed after functional test was completed.The sample passed functionality and performance testing.The root cause of the customer's complaint could not be established; the returned sample met specifications.After investigation of this complaint, it has been determined that this sample met specifications; therefore, no corrective action is required at this time.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key9528482
MDR Text Key177843502
Report Number2028159-2019-02327
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP-JAPAN
Device Catalogue Number8065751817
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2020
Date Manufacturer Received04/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CONSTELLATION SURGICAL PROCEDURE PAK
-
-