It was reported by the rn that the intra-aortic balloon pump (iabp) had stop pumping and the alarm was, purge failure.
The rn called the hotline and reported to the clinical support specialist (css) that the pump alarmed for purge failure.
There is both electrocardiogram (ecg) and arterial pressure (ap) waveform on the pump.
The gas driveline tubing was secure and there is adequate helium in the tank.
The staff reset the alarm and the pump started pumping again.
Again, the purge failure alarm occurred.
As a result, the pump was powered off and back on.
The pump is now pumping without alarming.
This took less than 5 seconds and the patient was stable throughout.
There was no report of patient injury or consequence.
|