• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1650DE
Device Problem Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/16/2019
Event Type  malfunction  
Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the battery cord was not working.The battery was exchanged.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the battery was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Failure analysis of the returned battery revealed a missing locking mechanism in the connector of the battery.Functional testing revealed that the battery was unable to have a secure connection while charging or when providing power to the controller.As a result, the reported event was confirmed.The most likely root cause of the reported event can be attributed to wear and/or the handling of the battery.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
This regulatory report is being submitted as part of a retrospective review and remediation per d00595828 due to an fda audit observation.Corrected fields: b5 event description was updated to include previously left out non-reportable information.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the battery would not charge and the cord was not working.The battery was exchanged.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY
Type of Device
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9530160
MDR Text Key189175832
Report Number3007042319-2019-14655
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00888707005708
UDI-Public00888707005708
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/31/2019
Device Model Number1650DE
Device Catalogue Number1650DE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/17/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
1103 VAD
Patient Age73 YR
Patient SexMale
-
-