Model Number 1012455-12
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/04/2019
Event Type  malfunction  
Manufacturer Narrative
Exemption number e2019001. The device was not returned for evaluation. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event. Additionally, a review of the complaint history identified no other similar incidents from this lot. The investigation determined the reported complaint appears to be related to operational context. There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that the procedure was performed to treat a moderately calcified, mildly tortuous de novo proximal right coronary artery that was 90% stenosed. Pre-dilatation was performed with an unspecified balloon. A 4. 0x23mmxience sierra stent was deployed and despite full dilatation of the balloon, it was noted via optical coherence tomography (oct) imaging that the stent had poorly apposed to the vessel wall as the vessel was wide. Therefore, a 5. 0x12mm rx nc trek balloon dilatation catheter was attempted to be used for post-dilatation, but the balloon ruptured at 6 atmospheres. The balloon catheter was replaced with a same size non-abbott balloon to successfully complete the procedure. Oct imaging confirmed the stent had fully apposed to the wall. There were no adverse patient effects and no clinically significant delay in the procedure. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNC TREK RX
Type of DeviceCORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key9531660
MDR Text Key178161452
Report Number2024168-2019-15062
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K103153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial,Followup
Report Date 01/29/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number1012455-12
Device Catalogue Number1012455-12
Device Lot Number90715G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/17/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/15/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-