Model Number 283512
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Event Description
It was reported by the affiliate that during an unknown procedure the cable of the micro tornado hp w handcontrol was damaged and it did not old burs.The procedure was completed using a replacement.No patient consequence and no surgical delay was reported.No additional information was provided.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary
=
> the complaint device was received at the service center and evaluated.It was reported that the cable of the microtornado hp handcontrol was damaged and did not hold burs.Per service manual operational and diagnostic, the reported failure was confirmed.During evaluation, the motor cable was found damaged since the cables were exposed; therefore, it was replaced.The insulation was damaged too near the bend protection.In addition, the values of the resistance were out of the range, in consequence the hand control set was replaced.Finally, the buttons were found not functioning, thus the service was completed replacing the buttons.The repair and testing of the unit was completed per the service manual, bringing the unit back to full functionality.The unit passed all functional tests and is fully operational.The possible root cause for the reported failure can be attributed to the damage the unit had which may cause that the unit stopped to working.Also, the short circuit on the motor cable may be as a failure to electrical components, and the buttons failure can be related to lack of maintenance.However, this cannot be conclusively affirmed.As a potential cause cannot be associated to manufacturing, therefore a manufacturing record evaluation is not required.At this time, no corrective action is required, and no further action is warranted, as the device was repaired and is fully functional.However, depuy synthes mitek will continue to track any related complaints within this device family to monitor the extent to which this complaint is observed in the field.Udi: (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d10, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO TORNADO HP W HANDCONTROL
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9532336
MDR Text Key199310117
Report Number1221934-2019-60207
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Type of Report Initial,Followup,Followup
Report Date 12/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2019
Date Manufacturer Received03/13/2020
Patient Sequence Number1
-
-