• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72200616
Device Problem Overheating of Device (1437)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Event Description
It was reported that the motor drive unit became hot.As the allegation was noticed before surgery, the patient was not involved at the time.Injuries were not stated.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the exterior of product and no physical damage was observed.Mdu failed functional tests with hand piece overload when power cord was bent or twisted.Power cord assembly grew warm during testing.After troubleshooting, the cause of the overheating was observed to be a defective power cord assembly.A visual inspection was performed on the power cords external covering and no physical damage was observed.It was determined that the power cord has shorted and/or opens internal wiring.Motor and hall board were tested and passed functional testing.The complaint investigation has concluded that this unit has succumbed to physical damage to power cord.Factors which can contribute to a shorted power cord include rolling a heavy cart over the cord or excessive bending of the cord.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures; this failure mode will be trended by post market surveillance to assess for any necessary corrective actions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT HAND CNTRL PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key9533816
MDR Text Key173342552
Report Number1643264-2019-00922
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010555472
UDI-Public03596010555472
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 02/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200616
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2019
Date Manufacturer Received01/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-