• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Catalog Number 400SMTHXSFT0306
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/10/2019
Event Type  malfunction  
Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
The patient was undergoing a coil embolization procedure in the anterior cerebral artery (aca) using penumbra smart coils (smart coils) and a non-penumbra microcatheter.During the procedure, the physician encountered resistance while advancing a smart coil to the target location and decided to retract the coil.On the third attempt, the physician encountered significant resistance and subsequently, the smart coil jumped back while the microcatheter shot forward and perforated the aneurysm wall.The smart coil was then completely removed.The procedure was completed using non-penumbra coils with the same microcatheter to stop the bleed and thrombose the remainder of the vessel.The patient was stable and was set to discharge home on (b)(6) 2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Type of Device
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key9536315
MDR Text Key177216082
Report Number3005168196-2019-02443
Device Sequence Number1
Product Code HCG
UDI-Device Identifier00814548015781
UDI-Public00814548015781
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K160832
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,12/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/08/2023
Device Catalogue Number400SMTHXSFT0306
Device Lot NumberF86345
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received12/10/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/09/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age71 YR
-
-