The patient was undergoing a coil embolization procedure in the internal carotid artery (ica) using penumbra smart coils (smart coils).During the procedure, a smart coil would not advance at all within its introducer sheath; therefore, it was removed.The procedure was completed using another smart coil.There was no report of an adverse effect to the patient.
|
Results: the pet lock was intact on the proximal end of the pusher assembly.The pusher assembly was kinked at approximately 138.0 cm from the proximal end.The pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.The embolization coil was undamaged and intact with the pusher assembly.Conclusions: evaluation of the returned smart coil revealed that the pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.If the mid-joint was retracted proximal to the friction lock, resistance will likely be experienced while attempting to advance the device through its introducer sheath.Subsequently, forcefully advancing the device against this resistance may have contributed to the pusher assembly kink, which was observed near the pusher assembly mid-joint.During functional testing, resistance was encountered as the pusher assembly mid-joint was advanced through the introducer sheath friction lock.After mid-joint was passed through the friction lock, the device was able to advance through its introducer sheath and a demonstration microcatheter with resistance due to the pusher assembly kink.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|