• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC KYPHX HV-R BONE CEMENT; CEMENT, BONE, VERTEBROPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC KYPHX HV-R BONE CEMENT; CEMENT, BONE, VERTEBROPLASTY Back to Search Results
Catalog Number C01A
Device Problem Migration (4003)
Patient Problem Pulmonary Embolism (1498)
Event Date 12/04/2019
Event Type  malfunction  
Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient underwent balloon kyphoplasty due to primary osteoporosis.Intra-operatively, during the operation the cement came into the vein on the side of the vertebral body.After that, the cement solidified.It was confirmed that the cement which came into the vein also solidified.The following day, it was checked when taking ct that the cement was not on the side of the vertebral body but came to the lungs, and has turned to pulmonary embolism.But there were no symptoms.There was no delay in overall procedure time as a result of this event.There were no patient complications as the result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHX HV-R BONE CEMENT
Type of Device
CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
stacie ziemba
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key9537591
MDR Text Key179246961
Report Number1030489-2020-00003
Device Sequence Number1
Product Code NDN
UDI-Device Identifier00643169097865
UDI-Public00643169097865
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K041584
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC01A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/05/2019
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-