• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR STEERABLE GUIDE CATHETER; VALVE REPAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR STEERABLE GUIDE CATHETER; VALVE REPAIR Back to Search Results
Catalog Number SGC0301
Device Problem Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 12/11/2019
Event Type  malfunction  
Manufacturer Narrative
Exemption number e2019001.The device was reported to be discarded.Investigation is not yet completed.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report during preparation the steerable guide catheter a leak was noted.It was reported that this was a mitraclip procedure to treat mitral regurgitation.During preparation of the steerable guide catheter (sgc), the sgc failed to hold column.The sgc was not used and never entered the patient anatomy.There was no patient involvement and no clinically significant delay in the procedure reported due to the device.No additional information was provided.
 
Manufacturer Narrative
G9: exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history did not identify any similar incidents.All available information was investigated and a definitive cause for the reported leak (loss of fluid column) in this incident could not be determined.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEERABLE GUIDE CATHETER
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9539075
MDR Text Key178156549
Report Number2024168-2020-00047
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648216824
UDI-Public08717648216824
Combination Product (y/n)N
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/06/2020
Device Catalogue NumberSGC0301
Device Lot Number90906U221
Was Device Available for Evaluation? No
Date Manufacturer Received01/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-