Catalog Number 190371
Device Problems Mechanical Problem (1384); Reflux within Device (1522)
Patient Problem No Patient Involvement (2645)
Event Date 12/16/2019
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process. A supplemental mdr will be submitted upon completion of this activity.
 
Event Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008k hemodialysis (hd) machine had a saline bag backfill. The biomed stated that a staff member visually observed the saline bag refilling with dialysate during the setup for a treatment. The biomed thought the blood pump rotor was causing the issue, so he replaced the rotor and returned the machine to service. There have been no reoccurrences of the reported event and the biomed believes the issue has been resolved. The biomed confirmed there was no patient involvement as the incident occurred during setup. The biomed discarded the rotor that was replaced; no sample available to be returned for evaluation. Multiple attempts were made to contact the user facility for additional information, and no further details could be obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name2008K HEMODIALYSIS SYS. OLC/DIASAFE PLUS
Type of DeviceDIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
matthew amaral
920 winter st
waltham, MA 02451
7816999758
MDR Report Key9547708
MDR Text Key181545936
Report Number2937457-2020-00017
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K994267
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/15/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number190371
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Device Age MO
Event Location No Information
Date Manufacturer Received01/07/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/28/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-