• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI ATTAIN PERFORMA MRI SURESCAN; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI ATTAIN PERFORMA MRI SURESCAN; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE Back to Search Results
Model Number 429888
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Endocarditis (1834); Fever (1858); Unspecified Infection (1930); Pain (1994); Thrombosis (2100); Fibrosis (3167)
Event Date 10/01/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient developed an infection of the endocardium with small vegetations in the tricuspid and mitral valves.It was noted that the bacteremia was caused by gram-positive bacteria.Preoperative diagnosis revealed evidence of venous thrombosis and fibrosis.It was also noted that the patient initially complained of back pain and a possible respiratory infection about a month ago with an intermittent fever over the last week.Rounds of antibiotics was administered with no signs of relief in symptoms.The cardiac resynchronization therapy defibrillator (crt-d) system was removed.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the full lead was returned and analyzed.No anomalies were found.Visual analysis of the lead indicated apparent explant damage.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN PERFORMA MRI SURESCAN
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9548978
MDR Text Key173656269
Report Number2649622-2020-00227
Device Sequence Number1
Product Code OJX
UDI-Device Identifier00643169002715
UDI-Public00643169002715
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date04/16/2017
Device Model Number429888
Device Catalogue Number429888
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/24/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age86 YR
-
-