The siemens customer service engineer (cse) performed a total service call, verifying all probe adjustment and sample probe to cuvette bottom calibrations.
The technical application specialist (tas) worked with the technician to prepare and run new fresh calibrators and quality controla for the ca19-9 assay.
The results were in specification.
The tas ran a study on a follow up call.
The mcm (master curve material) was running low but in range.
The cse verified and calibrated reagent probes, replaced ancillary probe and calibrated/adjusted all positions; calibrated ancillary que , primed the all lines, verified acid and base dispense volumes, and primed the lines.
Post service, the customer ran the quality control, no issues, level 1 and level 3 were in range.
The tas ran a study with the mcm and dilution.
The study passed.
Siemens healthcare diagnostics is investigating and has requested additional information from the customer.
The interpretation of results section of the instructions for use states: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings.
" the limitations section of the instructions for use states: warning: "do not use the advia centaur ca 19-9 assay as a screening test or for diagnosis.
Do not predict disease recurrence solely on levels of advia centaur ca 19-9.
Normal levels of advia centaur ca 19-9 do not always preclude the presence of disease.
Note: do not interpret serum levels of ca 19-9 as absolute evidence of the presence or the absence of malignant disease.
Before treatment, patients with confirmed gi carcinoma frequently have levels of ca 19-9 within the range observed in healthy individuals.
Additionally, elevated levels of ca 19-9 can be observed in patients with nonmalignant diseases.
Measurements of ca 19-9 should always be used in conjunction with other diagnostic procedures, including information from the patient's clinical evaluation.
The concentration of ca 19-9 in a given specimen determined with assays from different manufacturers can vary because of differences in assay methods, calibration, and reagent specificity.
Ca 19-9 determined with different manufacturers' assays will vary depending on the method of standardization and antibody specificity.
Therefore, it is important to use assayspecific values to evaluate quality control results.
".
|