MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC MAZOR RENAISSANCE SYSTEM INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC MAZOR RENAISSANCE SYSTEM INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number TPL0038
Device Problem Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/20/2019
Event Type  malfunction  
Manufacturer Narrative
Through log and export analysis, the investigating team has concluded the root cause of the registration failure was due to low image match between the ct and the fluoro, caused by the low-quality images acquired. Added hardware was also a contributing factor which influences the image match as well. If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a guidance system being used during a spinal procedure to treat degenerative discs with fixation. It was reported that procedure was planned for pedicle screw fixation at l4-l5 with l4-l5 alif interbody. The surgical team could not achieve successful registration labeling off of l3 with automatic registration. They attempted to adjust segmentation lines but still failed to achieve registration. Use of the surgical system was aborted. No patient harm was reported and surgery was delayed less than an hour.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMAZOR RENAISSANCE SYSTEM
Type of DeviceINSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key9575389
MDR Text Key184347991
Report Number1723170-2020-00106
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial
Report Date 01/10/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberTPL0038
Device Catalogue NumberTPL0038
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/22/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-