Stryker mako leg holders.
Numerous pieces have exposed fibers, two pieces have stress fractures at the point where the metal fitting is inserted.
Report of "someone getting stuck" by an exposed fiber.
Case type: tka.
Patient was under anesthesia.
Update : how many surgeries were involved with the reported issue? approximately 300 procedures to get to this point.
All the reported devices that are included in this complaint were for one or many surgeries? many ¿ spd called us concerned about exposed fibers and cracks.
Have you reported each incident separately that has the issues you have mentioned in this complaint? (if no, please submit a complaint for each issue.
) no ¿ multiple leg holders are exhibiting wear.
The items listed in the per either have exposed fibers or cracks.
Please confirm if there was a patient that was ¿stuck¿ by the exposed carbon fiber? not a patient ¿ a staff member.
|