• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED STERILIZATION PRODUCTS STERRAD 100NX STERILIZER ALLCLEAR 1-DR DUO; STERRAD EQUIPMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED STERILIZATION PRODUCTS STERRAD 100NX STERILIZER ALLCLEAR 1-DR DUO; STERRAD EQUIPMENT Back to Search Results
Model Number 10104-007
Health Effect - Clinical Codes Headache (1880); No Code Available (3191)
Date of Event 12/19/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).A field service engineer was dispatched to the customer site.The adapter converter, catalytic converter, oil mist filter, and vacuum pump oil were replaced to resolve the smoke/haze issue.Unit meets specifications and was returned to service.Asp complaint ref #: (b)(4).
 
Event or Problem Description
A customer reported an event of an odorless vapor/smoke emitting from the sterrad® 100nx sterilizer and one healthcare worker (hcw) experienced irritation, burning, and itchiness in the eye.Upon follow-up, the hcw noted he took his contacts off to rinse and flush his eyes, and the reaction lasted 4 hours.The hcw also had a ¿light headache¿ which lasted 1 hour.However, the hcw did not receive medical attention/treatment for his symptoms.The customer was advised to clear the facility and shut down the unit until serviced.An asp field service engineer was dispatched to assess the unit onsite.The injuries reported were not considered serious as the symptoms resolved on their own without any medical treatment.However, this event is being reported as a malfunction subsequent to a serious injury.
 
Additional Manufacturer Narrative
Asp investigation summary: the investigation included a review of the device history record (dhr), trending analysis of the smoke/haze issue, and system risk analysis (sra).The dhr was reviewed and no issues relating the failure mode were noted.The involved unit met manufacturer specifications at the time of release.Trending analysis of the smoke/haze issue was reviewed within the past six months and no significant trend was observed.The sra shows the risk for exposure to toxic or corrosive material to be "low." the parts were not available for return and further analysis.The assignable cause of the smoke/haze issue is the adapter converter, catalytic converter, oil mist filter, and vacuum pump oil.The field service engineer replaced these parts and confirmed the sterrad® 100nx was restored to proper function after service.The issue was resolved at the customer facility.Asp will continue to track and trend this issue.Asp complaint ref #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERRAD 100NX STERILIZER ALLCLEAR 1-DR DUO
Common Device Name
STERRAD EQUIPMENT
Manufacturer (Section D)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA
MDR Report Key9578279
Report Number2084725-2019-00010
Device Sequence Number18969577
Product Code MLR
UDI-Device Identifier10705037048694
UDI-Public10705037048694
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K071385
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10104-007
Device Catalogue Number10104007
Was the Report Sent to FDA? Yes
Initial Date Received by Manufacturer 12/20/2019
Supplement Date Received by Manufacturer07/10/2020
Initial Report FDA Received Date01/10/2020
Supplement Report FDA Received Date07/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-