Model Number 201-90411
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/16/2019
Event Type  Injury  
Manufacturer Narrative
This event was also reported against the centrimag motor under mfr. Report #2916596-2020-00002. Patient info was requested but not provided. No further information was provided. A supplemental report will be submitted when the manufacturer¿s investigation is completed.
 
Event Description
It was reported that the patient was on bi-vad centrimag (cmag) support with complaint of no flow showing on the right ventricular assist device (rvad) side. Patient was asymptomatic and flow appeared normal as patient was stable. Attempted new flow probe on the right side to no avail. Used the flow probe from the left side to ensure the flow probe worked and it did. That same flow probe was then attached to the rvad console and did not record a flow. The team then decided to switch out the rvad console with the backup console and flow again was recording. When the original rvad was attached to the mock loop, flow was normal. No further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCENTRIMAG 2ND GENERATION PRIMARY CONSOLE
Type of DeviceCONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ CH-8005
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key9582048
MDR Text Key175023512
Report Number2916596-2019-06108
Device Sequence Number1
Product Code DWA
UDI-Device Identifier07640135140702
UDI-Public07640135140702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131179
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 04/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model Number201-90411
Device Catalogue Number201-90411
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received04/03/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/05/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 01/13/2020 Patient Sequence Number: 1
-
-