MAUDE Adverse Event Report: BIOMERIEUX ITALIA S.P.A. VIDAS® ANALYSER - 99735

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMERIEUX ITALIA S.P.A. VIDAS® ANALYSER - 99735 Back to Search Results
Model Number 99735
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
A customer in the united states contacted biomérieux to report a qcv-detected failure of the vidas® analyzer (reference (b)(4), serial (b)(4)).The qcv detected failure occurred on (b)(6) 2019 the most recent successful qcv occurred on (b)(6) 2019.The customer performed a retrospective analysis following a qcv detected failure on position a1 for the affected timeframe.The customer stated a total of forty-seven (47) patients samples were tested at position a1 during the affected timeframe, six (6) patients samples were found to be underestimated for procalcitonin (pct).Results for sample c are listed below: sample c: pct reported value = 0.15 ng/ml ; pct repeat value = 21.06 ng/mll.The customer reported that these results (original and corrected results) have been reported to the physician.There is no indication or report from the laboratory that the discrepant result led to any adverse event related to the patient's state of health.A biomérieux field service engineer (fse) visited the customer and confirmed the pump at position a1 was clogged.The fse cleaned the pump and replaced the seals.The fse then performed a qcv and leak test, all results passed.A biomérieux internal investigation will be conducted.
 
Manufacturer Narrative
An internal investigation was initiated following the notification of six falsely underestimated vidas® procalcitonin (pct) results identified following a qcv-detected failure in section a1 of the vidas® analyzer.It should be noted a qcv failure is not an abnormal behavior.It means that the qcv played its role as a functional control.This control is meant to detect residual risks, that are rare and sudden.Those risks are already present and accepted into the vidas system risk analysis.The local field service engineer (fse) visited the customer site on 30-nov-2019 to repair and qualify the instrument.The fse performed the following actions: visual inspection of the seals.Replacement of the seals.Visual inspection in the tray, on the door, on the shield insulate plate, and on the bottom of the frame.Performed a pump tester, which failed in section a1.Performed a pump cleaning on the section a.Performed a new pump tester after the cleaning: all results passed.Door plunger ball check.Stricker plate check.Spring integrity check.Free and smooth vertical movement of spr blocks.Cleaned and lubricated the screws holding the spr block.Checked spr block and adjusted.Checked tower height and adjusted.Checked pump block and adjusted.Checked retainer plate integrity.Checked try depth and adjusted.Performed a leak test and a qcv test to qualify the instrument.All results passed.Root cause: pump clog on position a1.The cause of the qcv failure was a clog in position a1.The problem was solved after cleaning the pump and the changing the seals.The system is now operating per manufacturing specifications.Corrected date note: the initial report was submitted with a contact date of 01-dec-2019.Further review of this event confirmed the event did not meet pre criteria until 10-dec-2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDAS® ANALYSER - 99735
Type of Device
VIDAS® ANALYSER - 99735
Manufacturer (Section D)
BIOMERIEUX ITALIA S.P.A.
via di campigliano 58 / loc. p
firenze 50012
IT  50012
MDR Report Key9582283
MDR Text Key220792885
Report Number9615037-2020-00004
Device Sequence Number1
Product Code DEW
UDI-Device Identifier03573026140427
UDI-Public03573026140427
Combination Product (y/n)N
PMA/PMN Number
K891385
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 02/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number99735
Was Device Available for Evaluation? No
Date Manufacturer Received01/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-