MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-751LNAL; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-751LNAL
Device Problem Device Alarm System (1012)
Patient Problems Hyperglycemia (1905); Renal Failure (2041)
Event Date 02/25/2019
Event Type  malfunction  
Manufacturer Narrative
Unit received alarm during basic occlusion test due to faulty fsr (gold).Unable to perform basic occlusion test, occlusion test, prime test, excessive no delivery test due to alarm.However, unit passed rewind test and displacement test.
 
Event Description
Customer reported via phone call that the insulin pump does not notify when insulin was not being delivered and resulted a high blood glucose value.Customer's blood glucose value was over 200 mg/dl.Insulin pump was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751LNAL
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key9582531
MDR Text Key174951862
Report Number3004209178-2020-54914
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169507760
UDI-Public(01)00643169507760
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 01/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-751LNAL
Device Catalogue NumberMMT-751LNAL
Device Lot NumberA4751LNALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2019
Date Manufacturer Received01/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/15/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age52 YR
-
-