MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC MAZOR X STEALTH EDITION ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC MAZOR X STEALTH EDITION ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number TPL0059
Device Problem Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/16/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a guidance system being used during a spinal procedure to treat degenerative discs using fixation. It was reported that confirmation images were taken after all screws were placed. L3 on the left was the first trajectory and only one that was according to plan. The s1 screws were acceptable, but superior compared to plan. All other trajectories were completely off and mainly superior, pointing through the endplates. Navigation was accurate during the entire case. The representative mentioned there were no issues during setup, the 10 point test, 3 point test, 3define scan, or registration. There was no patient harm and the procedure was not delayed more than an hour.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMAZOR X STEALTH EDITION
Type of DeviceORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key9582853
MDR Text Key175391571
Report Number1723170-2020-00130
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182077
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 01/27/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberTPL0059
Device Catalogue NumberTPL0059
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/27/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No