Model Number 4FC12
Device Problem Gas/Air Leak (2946)
Patient Problems Embolism (1829); Paralysis (1997)
Event Date 12/17/2019
Event Type  Injury  
Manufacturer Narrative
Product event summary: the data files were returned and analyzed. The files showed at least twelve applications were performed with catheter afapro28 with lot number 28147 on the event date. The flow rate and the pressure are within specification. There were three applications without ablation. Clinical issues (paralysis, anorexia, embolism) were encountered post procedure. In conclusion, there is no indication of a malfunction of the sheath related to the adverse event. If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that following a successful cryo ablation procedure which was touched up with radiofrequency, the day after the procedure, the patient's right hand did not move and they developed anorexia. A magnetic resonance imaging (mri) showed embolus-like silhouettes and air ingress was alleged. The patient's symptoms recovered. No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFLEXCATH ADVANCE STEERABLE SHEATH
Type of DeviceCATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA H9R 5Z8
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9591468
MDR Text Key175058174
Report Number3002648230-2020-00037
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 03/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number12342
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/19/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 01/15/2020 Patient Sequence Number: 1
Treatment
AFAPRO28 BALLOON CATHETER
-
-