Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Injury (2348)
Event Date 12/19/2019
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the evolution transfer carriage and found the lock pin springs were not operating properly.The lock pins are located on the bottom of legs of the transfer carriage and allow the wheels to lock into place.The root cause of the reported event can be attributed to the springs within the lock pins.The springs had worn causing the lock pins not to stay in place and for the transfer carriage to retract towards the employee resulting in the reported event.While onsite, the technician replaced the springs, tested the unit, and found the transfer carriage to be operating according to specification.No additional issues have been reported.
 
Event Description
The user facility personnel reported an employee injured their legs while operating their evolution transfer carriage.Medical treatment was sought but was not administered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUTION TRANSFER CARRIAGE
Type of Device
TRANSFER CARRIAGE
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key9592006
MDR Text Key199553411
Report Number3005899764-2020-00005
Device Sequence Number1
Product Code FLE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/19/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-